*Price of the test:
$56 per analysis
Acceptable Matrices:
Food Products, Food Ingredients, Nutraceuticals, Non-alcoholic Beverages
Unacceptable Matrices:
This method is not applicable to materials containing: silica, insoluble aluminum, tin, shellfish and oysters.
Some classes of compounds are unsuitable for closed vessel microwave digestion as they are highly reactive with nitric acid and other oxidizers which may become nitrated and potentially explosive. These include: Animal fat (lard), glycerin and glycols, volatile flavoring agents, filter paper, packaging material, and alcoholic beverages.
Limit of Quantitation:
100 ng/g (ppb)
Sample Size Requirements:
10 grams
Equipment:
Inductively Coupled Plasma Mass Spectroscopy (ICP-MS)
Reportable Units:
ppb = ng/g
Turn Around Time:
7-9 Business Days
Is Rush Available?
Yes
Method Reference:
AOAC 993.14 and 2015.06
Method Description:
Samples are prepared by microwave closed vessel digestion with nitric acid. Solutions are infused into an ICP-MS instrument to perform identity and quantitation based on the linear response of the elements using calibration standards containing the elements with internal standards.
Information Required by Submitter:
Please supply estimates for solid matrices in ppb (ng/g)
Additional Information:
This test will only measure the soluble portion of Aluminum present in your sample. If your sample contains both soluble and insoluble aluminum, please contact customer service at 1-800-245-5615 or specify the need for both soluble and insoluble Aluminum on your sample submission form.
Contact Us:
If you have questions about our testing services, please contact us at +1 (800) 245-5615.
*Prices and offerings are subject to change without notice.